FASTCheck15 Failed Test Protocol

Your Test Failed? Don’t Worry We Have Your Back!

Follow these instructions to ensure your DUWL safety by treating and retesting your water.

Graphic of failed water test showing two red, vertical lines.

Failed Test Protocol

1. Do you have a Sterisil® R2A Waterline Test readily available?

YES – my Sterisil® R2A waterline test is readily available.

If the answer is YES, first retest with FASTCheck15™ test strip. Next, document your results here.

  1. If results PASS, Shock as CFU count may be close to 500 CFU/mL. The Sterisil Straw does not have to be removed when shocking with Citrisil Shock.
  2. If results FAIL, follow the instructions to conduct the Sterisil® R2A Waterline Test and prepare the sampling.
  3. After collecting samples for the Sterisil® R2A Waterline Test, shock immediately. The Sterisil Straw does not have to be removed when shocking with Citrisil Shock. You may resume patient care after shocking is complete.
  4. A follow-up FASTCheck15™ test strip should be conducted in 1 week to ensure shock was effective.
  5. Mail in your samples to the lab for R2A processing. R2A results will automatically be located in your My Solutions Center once results become available.
  6. Line cleaning with an approved DUWL cleaner is recommended. Do not use bleach. Remove Sterisil Straw when cleaning lines.
  7. Contact a Solmetex Water Specialist if continued failure occurs.

2. Log Your Results in My Solutions Center

3. Continue to Test Monthly

NO – I do not have my Sterisil® R2A waterline test.

If NO, retest with a FASTCheck15™ test strip and document results here.

  1. If results PASS, Shock as CFU count may be close to 500 CFU/mL. The Sterisil Straw does not have to be removed when shocking with Citrisil Shock.
  2. If results FAIL, shock immediately. The Sterisil Straw does not have to be removed when shocking with Citrisil Shock. You may resume patient care after shocking is complete.
  3. A follow-up FASTCheck15™ test strip should be conducted in 1 week to ensure shock was effective.
  4. A Sterisil R2A Waterline Test is recommended to determine more definite CFU count.
  5. Mail in your samples to the lab for R2A processing. R2A results will automatically be located in your My Solutions Center once results become available.
  6. Line cleaning with an approved DUWL cleaner is recommended. Do not use bleach. Remove Sterisil Straw when cleaning lines.
  7. Contact a Solmetex Water Specialist if continued failure occurs.

2. Log Your Results in My Solutions Center

3. Continue to Test Monthly

The Sterisil® SafeWater Solution

An orange box of Citrisil All-In-One Waterline Shock Tablets.

Citrisil Shock

  • Elimina las bacterias causantes del mal olor
  • Fórmula no oxidante que limpia las líneas de flotación de las unidades dentales y la superficie interior de los frascos dentales.
A blue box containing FastCheck15 Waterline Test Strips.

FASTCheck15Prueba de línea de flotación

• A 15-minute in-office waterline test.
• The fastest and most affordable way to monitor water quality for bacteria.

Prueba por correo Sterisil R2A

Prueba de línea de flotación Sterisil R2A

Sterisil R2A bacterial tests utilizes an EPA validated R2A plating method to grow and count HPC bacterial colonies in samples collected from dental unit waterlines.

The Sterisil® 90 or 365 Day All-in-One Straw – Distilled & Municipal.

Pajita Sterisil

  • Potente tratamiento de la línea de flotación mediante tecnología de iones de plata.
  • Eficacia 20 veces superior a la de las pajitas de yodo, garantizada durante 365 días
  • Tratamiento de choque inicial incorporado

The Sterisil SAFEWATER Protocol

Cuadro del protocolo SafeWater

¿Ha suspendido un examen?

It happens.
Here’s what to do:

First Do This:

Golpee y vuelva a probar utilizando el mismo método.

Failed Again?

We can help. Call us at:
(877) 207-1551

Troubleshooting

  • If no lines appear, test is invalid. Repeat testing instructions with a new test kit.
  • Make sure that each step is followed closely, and the timer is set to the exact times noted in the instructions.
  • Too much water in the vial can dilute the reagent preventing any lines from displaying.

Questions? Speak to one of our experts! Call (877) 207-1551